Javascript is either disabled or not supported by this browser. This page may not appear properly.
1200 Degree Celsius Vacuum Heater System Specification
Operating Temperature

Vacuum Environment

Temperature Stability

Heater Power

Heater Life

Heater Insulation Materials

Vacuum System

Water Cooling

Cooling Method

Temperature Monitoring

Safety

Interlock Protection

Data Storage

Data Monitoring

Physical Characteristics

Mobility

Facility Power

Noise Level

1200 Degrees Celsius (1300C Max due to melting temperature)
                                 of stainless steel and Nickel)
10E-3 Torr

Within 3 degrees at 1200C

150 watts at 120VAC, Single Phase

1000 hours at operating temperature and vacuum

Free of impurities (no adhesivies or binders).
   Surface Outgas rate less than 1E-8 Torr-cc/cm^2-sec
3.8 CFM 2-stage Mechanical Pump
   1/2 HP, 115VAC, 50/60Hz, Single Phase
Not required

Thermal Conduction and Thermal-Vacuum Radiaiton

Digital (Type "S" thermocouples) with set-point temperature control
    and over-temperature limit.
Complies with National and International safety standards

Heater power "OFF" should vacuum or temperature exceed
   pre-set values.  Manual push-button reset.
Pre-selected intervals into computer memory

Vacuum and Temperature visually at the system and remotely
   on a computer terminal
Compact in size (less than 2 feet square foot print)
   and less than 50 lbs (plus 55 lbs for mechanical vacuum pump)
Moveable on casters with lock
   Capable of passing through normal doorway openings
20A, 115VAC, 60 Hz, Single Phase

-48dB (approximate)